The JOMESA system for cleanliness analysis |
JOMESA Particle standard
Microscopic
counting of residue particles on filter membranes contains the following steps:
Scanning the complete filter with a scanning stage, particle detection and
counting grouped in size classes.
Measurement results therefore depend on many parameters and system settings, like:
Calibration of microscope, focusing, magnification settings, adjustment of
scanning stage, parameters for particle detection (thresholds), etc.
To check particle counting systems, filters taken from the production line are not suitable for the following reasons:
· Inaccurate reproducibility, because
filter transport can add or remove residues. No possibility of cleaning or conserving.
·
The actual particle sizes are
unknown or not well defined.
· Production filters contain particles with sizes close to class size boundaries.
The
JOMESA Particle standard contains defined particles according to the ISO
16232 standard.

The standard is designed such that all particle sizes are sufficiently far away
from size borders, thus enabling reproducible measurements.
It is targetted for filter diameters 47mm and 25mm.
The particle standard is available with a calibration certificate, documenting the trace-ability to national standards.